Product Wiki · Recalls · Health Canada — Medical Devices

Eclipse Treatment Planning System — recall 48424

Health Canada — Medical Devices · Canada · 2012-09-17 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
48424
Date of alert
2012-09-17
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Eclipse treatment planning system where an outdated assigned Hounsfield Unit [Hu] value can be used for dose calculation when the clock on the Eclipse Client workstation is not synchronized with the database / system server. When the Eclipse client workstation clock is running behind the clock on the database / system server, it is possible that a change to the assigned HU value for a structure does not invalidate the density image stored in the cache, and for the subsequent dose and Monitor Unit [MU] calculation to be based on the outdated HU assignment. The resultant MUs may be higher or lower than expected, dependent on the specific HU assignments. Treatment of the patient using these values can therefore lead to under or over dose.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/eclipse-treatment-planning-sys

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.