Dimension Products – Iron Calibrations — recall 48256
Health Canada — Medical Devices · Canada · 2013-01-07 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Complaints received regarding iron calibrations outside of acceptable Calibration Slope (M) values (0.97-1.03) when using Iron Calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints received.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.