Product Wiki · Recalls · Health Canada — Medical Devices

Stellar 100 and Stellar 150 (2019-12-05) — recall 48030

Health Canada — Medical Devices · Canada · 2019-12-20 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
48030
Date of alert
2019-12-20
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Resmed received a confirmed case of an alarm buzzer not working under the following circumstances: the device has a failed electronic component and, the device is stored without AC power connected for more than 36 hours leading to full depletion of the battery and, the device powers on automatically when connected to AC power without pressing the power switch. The identified issue is a combination of software and a component failure that may occur in a small number of devices. If a device with is issue is used in the condition described above, the audible alarms may not operate properly.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/stellar-100-and-stellar-150-20

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.