Health Canada — Medical Devices · Canada · 2017-09-11 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The investigation report (IR) for Synchromed II pump corrosion defines the problem statement as the "Degradation of Synchromed II motor components due to corrosion, wear, and a combination of corrosion and wear can result in temporary, intermittent, and/or permanent motor stall." The root cause of failure is identified as the depletion of lubricant at the shaft/jewel bearing interface which initiates mechanical wear on shaft ends, resulting in wear particles that can mix with the remaining lubricant creating residue. Shaft wear results in temporary or permanent motor stall.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.