Oasis Chest Drain — recall 44459

Health Canada — Medical Devices · Canada · 2017-02-09 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
44459
Date of alert
2017-02-09
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: It has been discovered through internal testing that there have been a small quantity of drains manufactured where the valve could potentially not re-seal properly after depressing the valve. If the patient is connected to suction on an oasis drain (with a filled air-leak monitor), a manual vent valve that will not re-seal is immediately obvious if continuous bubbling in the water seal chamber is observed.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/oasis-chest-drain

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.