Mammomat Inspiration System - Main Unit — recall 44409
Health Canada — Medical Devices · Canada · 2017-02-01 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Potential software issues on Mammomat Inspiration with system version VB30 and biopsy functionality. The system may freeze and the biopsy cannot be continued in the following situations: the compression force drops, e.g. due to patient movement, during the active targeting phase, or, during the targeting phase the inversion tool is used while the scout image is active. As a result of the system freeze, the biopsy may have to be repeated.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.