Product Wiki · Recalls · Health Canada — Medical Devices

MONACO (2016-08-18) — recall 44359

Health Canada — Medical Devices · Canada · 2016-08-26 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
44359
Date of alert
2016-08-26
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: In a specific workflow where contours are edited and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner. If the patient is not saved and recalled, in a subsequently created plan the bev will show the contours as still present while the transverse, sagittal and coronal views will show the contours as deleted. The total volume of the structure reported in the dvh and the structure control will not match. If the study set is saved and reopened before plan creation the total volume of the structure reported in the dvh and the structure control will be corrected. A similar workflow could trigger the defect if contour deletions and edits are performed while a plan is loaded.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/monaco-2016-08-18

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.