LAERDAL SUCTION UNIT, VARIOUS COMPONENTS — recall 44336
Health Canada — Medical Devices · Canada · 2016-08-15 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: AWARE OF A PROBLEM RELATED TO USE OF THE LAERDAL SUCTION UNIT (LSU) WHEN USED WITH THE LSU BATTERY THAT HAS BEEN STORED OR INSTALLED A LOW TEMPERATURES FOR A PROLONGED PERIOD OF TIME. THE LSU MAY SHUT OFF AFTER SOME SECONDS WHEN OPERATED AT 350 MMHG OR 500+ MMHG SETTINGS.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.