AW Volumeshare - Seno Advantage Software Option — recall 25742
Health Canada — Medical Devices · Canada · 2012-09-18 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: GE Healthcare has become aware that patient mismatch without a warning message could occur on Seno Advantage (SA) Workstations under the following circumstances: 1. New non-mammographic images for patient a are being received (for example, via the network) at the same time that mammographic images for patient a are displayed on the high-resolution black and white monitors. 2. The non-mammographic images are then selected and displayed on the color monitor. 3. Mammographic images from a different patient are then selected on the high-resolution black and white monitors.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.