Product Wiki · Recalls · Health Canada — Medical Devices

Acute Renal Dialysis Machine Aquarius System and with Platinum… — recall 25741

Health Canada — Medical Devices · Canada · 2012-01-20 · Type III

As described by Health Canada — Medical Devices: Acute Renal Dialysis Machine Aquarius System and with Platinum Software

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
25741
Date of alert
2012-01-20
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Baxter received reports of 8 cases of broken/damaged wheels on the Aquarius. The damage to the wheels occurred during the moving of the device. The damaged wheel may result in loss of stability of the device. In the worst case, the device can fall over and potentially harm a person.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/acute-renal-dialysis-machine-a

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.