A) Programmer Application Software 2868 B) Latitude Secure Server — recall 21688
Health Canada — Medical Devices · Canada · 2012-01-23 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: A measured shock lead impedance that is greater than 125 Ohms or less than 20 Ohms will generate an alert message and a red alert. However, the value is not displayed on screen. Because of that some physicians think that their leads are defective which is not necessarily the case.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.