Recall Z-3201-2018 — Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour Plus LINK 2.4 wireless blood glucose monitoring system (glucometer, test strips and control solution) is designed for self-testing by persons with diabetes and the monitoring of glucose co
U.S. FDA — Device Enforcement · United States · 2018-05-16 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.