Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-3065-2024 — Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;

U.S. FDA — Device Enforcement · United States · 2024-07-30 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-3065-2024
Date of alert
2024-07-30
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Siemens Healthcare Diagnostics, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
10445146
Reported risk: During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.