Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-2666-2020 — SafeStep Huber Needle Set Port Access Kit; Catalog Numbers PA-0029 (UDI 00801741085482), PA-0029YN (UDI 00801741085499), PA-0030 (UDI 00801741085505), PA-0030YN (UDI 00801741085512), PA-0031 (UDI 00801741085529), PA-0031YN (UDI 00801741085536), PA-0032 (UDI 00801741085543), PA-0032YN (UDI 0080174108

U.S. FDA — Device Enforcement · United States · 2020-06-19 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2666-2020
Date of alert
2020-06-19
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Becton Dickinson & Company
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00801741085482
Reported risk: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.