Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-2629-2025 — Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

U.S. FDA — Device Enforcement · Netherlands · 2025-09-03 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2629-2025
Date of alert
2025-09-03
Country / jurisdiction
Netherlands
Risk level
Class II
Risk type
recall
Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00884838059108
Reported risk: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.