Recall Z-2547-2014 — Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL HEAD 46MM DIA ENDO FEMORAL HEAD 47MM DIA ENDO FEMORAL HEAD 48MM DIA ENDO FEMORAL HEAD 49MM DIA ENDO FEMORAL HEA
U.S. FDA — Device Enforcement · United States · 2014-07-22 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.