Product Wiki · Recalls · U.S. FDA — Device Enforcement

AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and — recall Z-2539-2018

U.S. FDA — Device Enforcement · United States · 2018-05-02 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2539-2018
Date of alert
2018-05-02
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Ad-Tech Medical Instrument Corporation
Product category
device
Product type
Devices
Reported risk: The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.