Product Wiki · Recalls · U.S. FDA — Device Enforcement

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only. — recall Z-2435-2020

U.S. FDA — Device Enforcement · United States · 2019-01-15 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2435-2020
Date of alert
2019-01-15
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Westmed, Inc.
Product category
device
Product type
Devices
Reported risk: A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.