Recall Z-2362-2015 — Alaris Syringe Pump, Model No. 8110. Delivers fluids. Expansion of Affected Units: BOM Housing Assembly Service, Part No. 49000226, and Lower Housing/Carriage Block Assembly, Part No. 148188-100.

U.S. FDA — Device Enforcement · United States · 2015-07-21 · Class I

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2362-2015
Date of alert
2015-07-21
Country / jurisdiction
United States
Risk level
Class I
Risk type
recall
Brand
CareFusion 303, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
49000226
Reported risk: Channel Error code is displayed on the PC unit with an audio and visual alarm, and on the syringe module. After the error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is detached and reattached.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.