Product Wiki · Recalls · U.S. FDA — Device Enforcement

Safety monitor included with System 8000 Base The Sarns" Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional — recall Z-2348-2012

U.S. FDA — Device Enforcement · United States · 2009-08-12 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2348-2012
Date of alert
2009-08-12
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Terumo Cardiovascular Systems Corporation
Product category
device
Product type
Devices
Reported risk: Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.