Product Wiki · Recalls · U.S. FDA — Device Enforcement

Medtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b)… — recall Z-2159-2021

U.S. FDA — Device Enforcement · United States · 2021-07-09 · Class II

As described by U.S. FDA — Device Enforcement: Medtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2159-2021
Date of alert
2021-07-09
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Medtronic Neuromodulation
Product category
device
Product type
Devices
Recall status
Ongoing — the recall is still in progress
Initiated by
Voluntary: Firm initiated
Distribution
Austria, Belgium, Brazil, Canary Islands, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Spain, Sweden, Switzerland, United Kingdom
Barcode (EAN/GTIN)
00763000324353
Reported risk: Specific PINs of the Vectris SureScan MRI lead kits contain the incorrect lead electrode spacing information printed on the shelf box and the sterile pack labeling.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.