Recall Z-2159-2012 — Cobas KRAS Mutation Test for In Vitro Diagnostic Use Product Usage: Usage: The Cobas KRAS Mutation Test, for use with the Cobas 4800 System, is a real-time PCR test intended for the identification of mutations in codons 12, 13 and 61 of the KRAS gene in DNA derived from formalin-fixed paraffin-
U.S. FDA — Device Enforcement · United States · 2012-02-24 · Class III
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: During the real-time stability testing for the cobas KRAS Mutation Test on the commercial lot PO6778, it was identified that the kit was just within specifications for the codon 61 peak height ratio, and that the kit has trended towards the upper specification limit at the 6 month time point but did passed the 6 month stability testing. The kit is labeled for 9 months shelf life stability. The d
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.