Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-2157-2025 — Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250

U.S. FDA — Device Enforcement · United States · 2025-06-12 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2157-2025
Date of alert
2025-06-12
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Bridge to Life Ltd
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
10860001991205
Reported risk: The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.