As described by U.S. FDA — Device Enforcement: Pausch Uroview FD fluoroscopic X-ray system UDI number (01)04260491620040 Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatmen
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Terminated — the recall action is complete and the FDA has closed it
Initiated by
FDA Mandated
Distribution
Nationwide Distribution
Reported risk: During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.