Recall Z-2142-2023 — HAMILTON C3 Ventilator REF 160005 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates
U.S. FDA — Device Enforcement · United States · 2023-06-15 · Class I
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.