Product Wiki · Recalls · U.S. FDA — Device Enforcement

Aeos Robotic Digital Microscope, Product Code: PV010 — recall Z-2139-2024

U.S. FDA — Device Enforcement · United States · 2024-05-13 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2139-2024
Date of alert
2024-05-13
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Aesculap Inc
Product category
device
Product type
Devices
Recall status
Ongoing — the recall is still in progress
Initiated by
Voluntary: Firm initiated
Distribution
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
Barcode (EAN/GTIN)
04046955206742
Reported risk: Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.