Recall Z-2135-2013 — ADVIA Chemistry A1c_3 Calibrators, Catalog Number 10491408, IVD --- COMMON/USUAL NAME: ADVIA Chemistry A1c_3 Calibrators Shipping or unit package: Level 1: 2 x 0.5 g; Level 2: 2 x 0.5 g; Level 3: 2 x 0.5 g; Level 4: 2 x 0.5 g -- Siemens Healthcare Diagnostics, Inc. Tarrytown, NY 10591 --- System I

U.S. FDA — Device Enforcement · United States · 2013-07-30 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-2135-2013
Date of alert
2013-07-30
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Siemens Healthcare Diagnostics
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
10491408
Reported risk: Siemens confirmed that the ADVIA¿ Chemistry Systems HbA1c method exhibits a positive bias of up to 12% for patient and CAP survey samples due to over-recovery when using A1c_3 Calibrator lots 1MD014, 1BD063, and 2GD014. The established ranges of CAP Survey Samples GH2-04, GH2-05 and GH2-06 and IFCC sample recoveries use ¿ 7% from the established mean as their acceptance criteria. This is the criterion that was used during testing to confirm the customer complaints. Depending upon quality control limits this issue may not have been detected..
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.