Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1991-2020 — Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, Premium, Urine Meter, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code P4P16XTSD - Product Usage

U.S. FDA — Device Enforcement · United States · 2020-04-30 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1991-2020
Date of alert
2020-04-30
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Covidien Llc
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
1828205964
Reported risk: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.