Recall Z-1988-2014 — GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

U.S. FDA — Device Enforcement · United States · 2014-06-04 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1988-2014
Date of alert
2014-06-04
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
GE Healthcare, LLC
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
608263214
Reported risk: GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.