Recall Z-1987-2020 — Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
U.S. FDA — Device Enforcement · United States · 2020-04-30 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.