Product Wiki · Recalls · U.S. FDA — Device Enforcement

Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a… — recall Z-1986-2015

U.S. FDA — Device Enforcement · United States · 2015-04-09 · Class II

As described by U.S. FDA — Device Enforcement: Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a measuring exercise equipment device designed to provide active and passive torque resistance to patient motion during rehabilitation therapy. The Primus that includes the optional Chop / Lift Bar.

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1986-2015
Date of alert
2015-04-09
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
BTE Technologies, Inc.
Product category
device
Product type
Devices
Recall status
Terminated — the recall action is complete and the FDA has closed it
Initiated by
Voluntary: Firm initiated
Distribution
Worldwide Distribution - Nationwide Distribution and to the countries of : Austria, Belgium, Canada, France, Germany, Greece, Hong Kong, Italy, India, Japan, South Korea, Mexico, P.R. China, Poland, Republic of South Africa, Romania, Saudi Arabia, Singapore, Spain, Switzerland, Thailand, and United
Reported risk: On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift Bar could get accidently detached from the snap hook in specific use scenarios when the bar is rotated and twisted around the snap hook. The snap hook could open, resulting in possible injury to the user due to a fall.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.