Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1944-2020 — iChem¿VELOCITY Urine Chemistry System-

U.S. FDA — Device Enforcement · United States · 2020-03-30 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1944-2020
Date of alert
2020-03-30
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Beckman Coulter Inc.
Product category
device
Product type
Devices
Reported risk: A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.