Product Wiki · Recalls · U.S. FDA — Device Enforcement

bioMerieux Vitek 2 Systems Software version 6.01 software, used with… — recall Z-1908-2013

U.S. FDA — Device Enforcement · United States · 2013-06-14 · Class II

As described by U.S. FDA — Device Enforcement: bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT V2S 6.01ML1. Part number: 414537. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antim

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1908-2013
Date of alert
2013-06-14
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Biomerieux Inc
Product category
device
Product type
Devices
Recall status
Terminated — the recall action is complete and the FDA has closed it
Initiated by
Voluntary: Firm initiated
Distribution
Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Belarus, Benelux, Brazil, Bulgaria, Canada, Chile China, Costa Rica, Croatia, Czech Republic, Dominican Republic, Dutch Antilles, Ecuador, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, I
Reported risk: A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.