Recall Z-1796-2015 — ELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted.
U.S. FDA — Device Enforcement · United States · 2015-05-29 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Mortara Instrument, Inc. has recently become aware of a potential safety hazard involving our ELI 380 electrocardiograph. When used in a particular workflow, acquired ECG waveforms for one patient may become associated with the patient demographics for a different patient when the record is transmitted to a records management system.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.