Recall Z-1777-2019 — Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are han
U.S. FDA — Device Enforcement · United States · 2019-05-20 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.