Recall Z-1755-2024 — Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

U.S. FDA — Device Enforcement · United States · 2024-03-14 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1755-2024
Date of alert
2024-03-14
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Johnson & Johnson Surgical Vision, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
05050474573376
Reported risk: Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.