Recall Z-1720-2025 — Washer Disinfector Aquadis 56; Model Number: 56A

U.S. FDA — Device Enforcement · Sweden · 2025-04-02 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1720-2025
Date of alert
2025-04-02
Country / jurisdiction
Sweden
Risk level
Class II
Risk type
recall
Brand
Getinge Disinfection Ab
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
07340153710344
Reported risk: The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.