Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1642-2026 — Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is

U.S. FDA — Device Enforcement · United States · 2026-02-25 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1642-2026
Date of alert
2026-02-25
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Olympus Corporation of the Americas
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00821925036000
Reported risk: Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.