Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1616-2014 — ACCU-CHEK Compact Plus Test Strips For use with: ACCU-CHEK Compact Plus Meters; 17 count sample strip 51 count retail strip; 102 count retail strip; 51 count MedBen strip 51 count Mail Order strip; Sample Kit 16 per case. The ACCU-CHEK Compact Plus Test Strips are for use with the ACCU-CHEK Co

U.S. FDA — Device Enforcement · United States · 2014-04-14 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1616-2014
Date of alert
2014-04-14
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Roche Diagnostics Operations, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
5599423160
Reported risk: Roche Diabetes Care has become aware the ACCU-CHEK Compact Plus test strips may produce erroneously low blood glucose readings in patients undergoing Ceftriaxone therapy. The interference is not described in the product labeling.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.