Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1560-2013 — Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defined scripts and processes on a Freedom EVO platform (Touch Tools Suite is an optional software add-on for Freedom EVOware (the base operational software for the Freedom EVO 2 i

U.S. FDA — Device Enforcement · United States · 2013-02-19 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1560-2013
Date of alert
2013-02-19
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Tecan US, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
10641100
Reported risk: WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.