RetractaBed , Clinical Contour, Med - Mizer Inc..... Model SS 762 ( 76 " bed w/ 2 motors); 763 (3 motors); 763 A ( 3 motors, autocontour) 120 V, ...Max Safe Working Load 450 LBS. Acute/Longterm Care AC Powered Adjustable Bed for geriatric care. — recall Z-1541-2014

U.S. FDA — Device Enforcement · United States · 2014-04-18 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1541-2014
Date of alert
2014-04-18
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Med-Mizer, Inc.
Product category
device
Product type
Devices
Reported risk: All configurations of the SS Retractabed, Clinical Contour, made prior to June 7 2008 may have had insufficient welding on the foot cross tube. This tube is responsible for holding the two support rails together when using the foot side rails and keeps them from separating from the frame. If the weld on the tube breaks, it can lead to bed collapse or the rollers coming out of the channel on th
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.