Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1515-2022 — cobas e801 Immunoassay Analyzer

U.S. FDA — Device Enforcement · United States · 2022-07-14 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1515-2022
Date of alert
2022-07-14
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Roche Diagnostics Operations, Inc.
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
07613336158456
Reported risk: Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.