Recall Z-1476-2014 — FB Reagent (REF 70562). FB reagent is an additional test used for revealing the results of some miniaturized biochemical test included in the API and Rapid ID 32 strips. The API and ID 32 strips product lines are a standardized system combining several biochemical tests, which enables group or sp

U.S. FDA — Device Enforcement · France · 2014-03-06 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1476-2014
Date of alert
2014-03-06
Country / jurisdiction
France
Risk level
Class II
Risk type
recall
Brand
BioMerieux SA
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
1002717180
Reported risk: bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.