Recall Z-1376-2023 — Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) MNDR-600, Western Oxytote DTE Digital VIPR (head only) MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) MNDS-603,
U.S. FDA — Device Enforcement · United States · 2023-01-03 · Class II
⚙AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Reported risk: Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.