Reported risk: Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting per low-power field (/LPF)* or *per microliter. This may occur during initial method validation or if settings are altered after the initial validation.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.