Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1320-2024 — I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 80mm,Part Number S65 ST180

U.S. FDA — Device Enforcement · France · 2024-01-26 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1320-2024
Date of alert
2024-01-26
Country / jurisdiction
France
Risk level
Class II
Risk type
recall
Brand
In2Bones, SAS
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
3760225710692
Reported risk: This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.