Recall Z-1248-2019 — Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), product code 35700BAX - Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood,

U.S. FDA — Device Enforcement · United States · 2018-06-21 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1248-2019
Date of alert
2018-06-21
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Baxter Healthcare Corporation
Product category
device
Product type
Devices
Reported risk: The flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not identify issues with pumps that could cause inaccuracies in flow rates which, in turn, could lead to over- or under- infusion. Potential risk to the patient depends on a number of factors including treatment and patient comorbidities. Depending on these factors, the patient could experience serious adverse health consequences.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.