Product Wiki · Recalls · U.S. FDA — Device Enforcement

Recall Z-1151-2021 — The PrisMax System Version II Hemodialysis Delivery System: Automatic Repositioning System Pump Assembly, specifically tubing, Model Numbers: a) 955626 b) 955558 c) 955701 PrisMax V2 Preventive Maintenance Kit, Product Code SC616 ARPS PUMP, Product Code SC6035 ARPS Pump Assembly, Product Code

U.S. FDA — Device Enforcement · United States · 2021-01-07 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1151-2021
Date of alert
2021-01-07
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Baxter Healthcare Corporation
Product category
device
Product type
Devices
Barcode (EAN/GTIN)
00085412639499
Reported risk: Variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during system self-test and therapy. This tubing is normally replaced during annual preventative maintenance. Alarm Situations: During System Self-Test: Alarm code B1215 (ARPS Self-Test failure); During Therapy: B1273 (ARPS Leak), B1215 (ARPS Pressure self-test), B1116 (Return Pressure self-test Failure), B1539 (Access Pod Reposition Failure), B1541 (Effluent Pod Reposition Failure), B1540 (Effluent Filter Pod Reposition Failure). If an alarm occurs, it may lead to delay or interruption of therapy. In the event that therapy is terminated without returning blood to the patient, blood loss may occur.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.