A warming pad, a component of the PINTLER PEAK & PINTLER SUMMIT Heated OR pad. PPWS-001 indicated for use in any condition where patient warming is desired as determined by clinical personnel, in accordance with facility normo-thermia protocol — recall Z-1139-2019
U.S. FDA — Device Enforcement · United States · 2018-04-20 · Class II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.