Recall Z-1128-2013 — SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist

U.S. FDA — Device Enforcement · United States · 2012-12-13 · Class II

AI-assisted, human-edited — this page is drafted with AI from public-source data, then reviewed before publication. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-1128-2013
Date of alert
2012-12-13
Country / jurisdiction
United States
Risk level
Class II
Risk type
recall
Brand
Sr Instruments Inc
Product category
device
Product type
Devices
Reported risk: SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.
Primary source: https://www.fda.gov/safety/recalls-market-withdrawals-safety-alerts

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.